THEORY OF OPERATION
Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
60
10.4.
Electronic Operation
10.4.1.
Sensor Module
This Section replaces the
Sensor Module and Sample Chamber
section in the T101
Manual - PN 07266.
Electronically, the T102 sensor module is a group of components that: create the UV
light that initiates the fluorescence reaction between SO
2
and O
3
; sense the intensity of
that fluorescence and generate various
electronic signals needed by the analyzer to
determine the SO
2
concentration of the sample gas (see Section 9.1) and sense and
control key environmental conditions such as the temperature of the sample gas and the
PMT.
Figure 10-2. T102 Sensor Module
These components are divided into two significant subassemblies. The sample chamber
and the PMT assembly.
•
Figure 9-3 shows an exploded view of the sample chamber assembly
•
Figure 9-5 shows an exploded view of the PMT Assembly
07267B DCN6485
Содержание 501 TRS
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Страница 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Страница 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Страница 54: ...07267B DCN6485 ...
Страница 90: ...TROUBLESHOOTING and REPAIR Teledyne API T102 501 TRS Addendum to T101 Operation Manual 88 07267B DCN6485 ...
Страница 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Страница 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Страница 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Страница 105: ...APPENDIX D Wire List and Electronic Schematics ...
Страница 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Страница 110: ...D 6 07267B DCN6485 ...
Страница 111: ......