Maintenance
3000TA- EU
Teledyne Analytical Instruments
70
Before replacing the cell:
a)
Check your span gas to make sure it is within
specifications.
b)
Check for leaks downstream from the cell, where oxygen
may be leaking into the system.
If there are no leaks and the span gas is OK, replace the cell. The
only way to avoid all fail messages is to turn the unit off and then back
on.
5.3.3 Removing the Micro-Fuel Cell
The Micro-Fuel cell is located inside the stainless steel cell block
behind the front panel (see Figure 5-1). To remove an existing cell:
1.
Remove power to the instrument by unplugging the power
cord at the power source.
2.
Open the inner door by unlatching it and swinging it open on
its hinges.
3.
Leave the cell block installed. Place one hand underneath the
cell block, ready to catch the cell holder and Micro-Fuel Cell.
4.
Lift up on the stainless steel gate in front of the cell block.
This releases the cell and cell holder from the block. The cell
and holder will fall into your hand.