Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
53
#### % Span
Slope=#### ppm/s
When the Span value displayed on the screen is sufficiently stable,
press
Enter
. (Generally, when the Span reading changes by 1% or less
of the full scale of the range being calibrated for a period of ten minutes
it is sufficiently stable.) Once
Enter
is pressed, the Span reading
changes to the correct value. The instrument then automatically enters
the Analyze function.
4.4.3 Span Failure
The analyzer checks the output of the cell at the end of the span. If
the raw output of the cell is less than 0.5 uA/ppm O
2
, the span will not
be accepted. The analyzer will return to the previous calibration values,
trigger the System Alarm, and display in the VFD:
Span Failed!!
This message will be shown for five seconds and the instrument
will return to the Analyze mode. In the upper right hand corner of the
VFD display “FCAL” will be shown. This message flag will help the
operator troubleshoot in case calibration was initiated remotely. To reset
the alarm and the flag message, the unit must be turned off by cycling
the standby key . It will reset if the next span cycle is correct.
A trace cell is unlikely to fail span. As explained before, when the
sensor reaches the end of its useful life, the zero offset begins to rise
until the analyzer finds the zero unsatisfactory. Nevertheless, feeding the
wrong span gas or electronics failure could set this feature off at the end
of the span. Consider this before replacing the cell.
4.5 The Alarms Function
The Model 3000TB is equipped with 2 fully adjustable concen-
tration alarms and a system failure alarm. Each alarm has a relay with a
set of form “C" contacts rated for 3 amperes resistive load at 250 V ac.
See Figure 3-6 in Chapter 3
Installation
and/or the Interconnection
Diagram included at the back of this manual for relay connections.
Содержание 3000TB
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Страница 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......