Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
59
interval (6 months is the default setting) after which a notification on the
Analyze
screen will advise that sensor maintenance is needed. This
function can be set so that in addition to the notification, it also triggers
the system failure alarm. It can also be turned off.
To access the Maintenance function use the Use the ◄ / ► arrow
keys to navigate to the fourth
System
menu.
Maintenance Serial#
Stream SerMode More
Use the RIGHT button to move the blinking to
Maintenance
and
then press
Enter
to go to the next screen.
Set: M: 0 D: 7 On
Nxt: M: 0 D: 6 Reset
The timer in the maintenance function is set in months (M) and
days (D) on the top line of the display. The maximum setting is M=30
and D=30. The default is M=6 and D=0. The user has the option of
turning the function
On
or
Off
or to have it trigger the system alarm
(
Alarm
).
To change the notification period, use the RIGHT arrow key to
move the blinking to the appropriate field M(onth) or D(ay) and then use
the ▲/▼ keys to increment or decrement the value. Press
Enter
when
the appropriate setting is displayed.
The second line is the actual countdown and displays the time left
before a notification or alarm will occur. It counts down until the
Maintenance warning is set. To re-start the counter the operator must
select “
Reset
” and press
Enter
. The operator is prompted to press
Enter
again to re-affirm the restart of the counter.
Reset Maint Timeout?
<ENT> Yes <ESC> No
Pressing ESC will return you to the previous screen.
Содержание 3000TB
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Страница 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......