Installation
Model 3000TB
Teledyne Analytical Instruments
32
Figure 3-10: Gas Connector Panel
SAMPLE IN:
In the standard model, gas connections are made at the
SAMPLE IN
and
EXHAUST OUT
connections. Calibration gases must be tee'd into
the Sample inlet with appropriate valves.
The gas pressure in should be reasonably regulated. Pressures
between 3 and 40 psig are acceptable as long as the pressure, once
established, will keep the front panel flowmeter reading in an acceptable
range (0.1 to 2.4 SLPM). Exact figures will depend on your process.
If greater flow is required for improved response time, install a
bypass in the sampling system upstream of the analyzer input.
Note: If the unit is for vacuum service, the above numbers apply
instead to the vacuum at the EXHAUST OUT connector,
described below, with minus signs before the pressure
readings.
ZERO IN and SPAN IN:
These are additional ports for inputting span gas and zero gas.
There are electrically operated valves inside for automatic switching