Installation
Model 3000TB
Teledyne Analytical Instruments
28
3.3.8 RS-232 Port
The digital signal output is a standard RS-232 serial communications
port used to connect the analyzer to a computer, terminal, or other digital
device. The pinouts are listed in Table 3-3.
Table 3-3: RS-232 Signals
RS-232 Sig
RS-232 Pin
Purpose
DCD
1
Data Carrier Detect
RD
2
Received Data
TD
3
Transmitted Data
DTR
4
Data Terminal Ready
COM
5
Common
DSR
6
Data Set Ready
RTS
7
Request to Send
CTS
8
Clear to Send
RI
9
Ring Indicator
The data sent is status information, in digital form, updated every
two seconds. Status is reported in the following order:
The concentration in ppm or percent
The range in use (HI, MED, LO)
The span of the range (0-100 ppm, etc)
Which alarms—if any—are disabled (AL–x DISABLED)
Which alarms—if any—are tripped (AL–x ON).
Three input functions using RS-232 have been implemented to
date. They are described in Table 3-4.
Table 3-4: Commands via RS-232 Input
Command
Description
as
<enter>
Immediately starts an autospan.
az
<enter>
Immediately starts an autozero.
st
<enter>
Toggling input. Stops/Starts any status
message output from the RS-232, until
st
<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 3-6 lists certain RS-232 values that are required by the 3000TB
implementation.
Содержание 3000TB
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Страница 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......