Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
57
Use the ◄ / ►arrow keys to blink the range to be set: low (
L
),
medium (
M
), or high (
H
).
Use the ▲/▼ arrow keys to enter the upper value of the range (all
ranges begin at 0 ppm). Repeat for each range you want to set. Press
Enter
to accept the values and return to
Analyze
mode. (See note below.)
Note: The ranges must be increasing from low to high, for
example, if range 1 is set as 0–100 ppm and range 2 is set
as 0–1,000 ppm, range 3 cannot be set as 0–500 ppm
since it is lower than range 2.
Ranges, alarms, and spans are always set in ppm units (over the
entire 0-250,000 ppm range), even though all concentration-data outputs
change from ppm units to percent when the concentration is above
10,000 ppm.
4.6.2 Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing
the analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, enter the range
function by pressing the
Range
button on the front panel.
Use the ◄ / ►arrow keys to move the blinking over
AUTO
.
Use the ▲/▼ arrow keys to switch from
AUTO
to
FX/LO, FX/M ED
,
or
FX/H I
to set the instrument on the desired fixed range (low, medium,
or high).
L—100 M—1000
H—1 0000 Mode—FX/ LO
or
L—100 M—1000
H—1 0000 Mode—FX/MED
or
L—100 M—1000
H—1 0000 Mode—FX/ HI
Press
Escape
to re-enter the
Analyze
mode using the fixed range.
Содержание 3000TB
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Страница 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......