Operation
3000TB
Teledyne Analytical Instruments
40
4.3.2 Setting up an Auto-Cal
When proper automatic valving is connected (see Chapter 3,
Installation
), the analyzer can cycle itself through a sequence of steps
that automatically zero and span the instrument.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000TB is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
To setup an
Auto–Cal
cycle:
Choose
System
from the Function buttons. The LCD will display
five subfunctions.
TRAK/HLD Auto—Cal
PSWD Logout More
Use
◄►
arrows to blink
Auto—Cal
, and press
Enter
. A new screen
for
Span/Zero
set appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Press
◄►
arrows to blink
Span
(or
Zero
), then press
Enter
again.
(You won’t be able to set
OFF
to
ON
if a zero interval is entered.)
A
Span Every
... (or
Zero Every
...) screen appears.
Span Every 0 d
Start 0 h from now
Use ▲/▼ arrows to set an interval value, then use ◄
►
arrows to
move to the start-time value. Use ▲/▼ arrows to set a start-time value.
To turn
ON
the
Span
and/or
Zero
cycles (to activate
Auto-Cal
):
Press
System
again, choose
Auto—Cal
,
and press
Enter
again. When
the
Span/ Zero
values screen appears, use the
◄►
arrows to blink the
Span
(or
Zero
)
OFF/ON
field. Use
▲/▼ arrows to set the
OFF/ON
field to
ON
. You can now turn these fields
ON
because there is a
nonzero span interval defined.
Содержание 3000TB
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 49: ...Installation Model 3000TB Teledyne Analytical Instruments 34 BLANK PAGE ...
Страница 81: ...Operation 3000TB Teledyne Analytical Instruments 66 BLANK PAGE ...
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......