sample preparatIon
A properly capped and sealed vial is critical in maintaining sample integrity during headspace autosampling. The
following are guidelines for proper vial preparation.
When a vial is undercrimped, the cap shape will be concave and the cap can be easily rotated. The “tuck-under” on the
vial will be at an angle, not in contact with the underside of the collar. If you feel some resistance when trying to rotate
the cap, it is probably sufficiently crimped.
Installation & Setup • Page 3-2
Figure 3-1: Properly Sealed Vial
Figure 3-2: Undercrimped Vial
basic operation
Содержание 15-0800-074
Страница 1: ...Versa User Manual www teledynetekmar com Versa Manual Part 15 0800 074 ...
Страница 8: ...Versa User Manual www teledynetekmar com Chapter 1 Introduction ...
Страница 21: ...Versa User Manual www teledynetekmar com Chapter 2 Installation Setup ...
Страница 36: ...Versa User Manual www teledynetekmar com Chapter 3 Basic Operation ...
Страница 44: ...Versa User Manual www teledynetekmar com Chapter 4 Versa TekLink 2G ...
Страница 61: ...Versa User Manual www teledynetekmar com Chapter 5 21 CFR Part 11 Compliance ...
Страница 70: ...Versa User Manual www teledynetekmar com Installation Setup Page 2 6 Chapter 6 Method Development ...
Страница 77: ...Versa User Manual www teledynetekmar com Chapter 7 Maintenance Troubleshooting ...
Страница 90: ...Versa User Manual www teledynetekmar com Chapter 8 Diagrams ...
Страница 91: ...Versa User Manual www teledynetekmar com Detailed Plumbing Diagrams Page 8 2 ...
Страница 100: ...Versa User Manual www teledynetekmar com Chapter 9 Index ...