2. Gas pressure at the source must be high enough to:
• Allow at least a 65 psi pressure drop at every pressure regulator.
• Travel the distance from the source to the Versa.
• Provide the required gas pressure at the Versa. Operation of the Versa requires
pressurize gas at an incoming (supply) pressure of 45 to 140 psig.
3. If the tubing that supplies the gas depends on the maximum pressure drop allowable for your setup:
• If the pressurize gas supply is close to the Versa, use the pre-installed 1/8” blue tubing.
• If any of the following circumstances are present, it may be necessary to reduce the pressure by using 1/4”
tubing to connect to the pre-installed 1/8” blue tubing:
a. Lengths greater than 15’.
b. A single source supplies several units.
c. A single source is subject to high demand for gas.
4. Make sure the gas supply tubing lengths are adequate. Be generous when cutting lengths of tubing for local
supply lines. A relatively long coil of tubing between the supply and the Versa allows you to move the unit without
disconnecting the plumbing. If necessary, additional tubing can be added to increase the length.
5. Make sure the gas line fittings and regulators are the correct size and type. Consult your local gas supplier for the
proper type and size of cylinder valves. Select compatible pressure regulators based on the required valves. Keep
these considerations in mind:
• Use good quality pressure regulators with stainless steel diaphragms. The regulators reduce the high source
pressures to the Versa. Teledyne Tekmar recommends using a single, two-stage regulator rather than two
single-stage pressure regulators to meet the Versa’s pressure specifications.
• On/Off valves, while not essential, are very useful when mounted on the outlet fitting of a two-stage regulator.
• If pipe thread connections are required in your gas supply lines, seal them with instrument-grade Teflon® tape.
mInImUm compUter reqUIrements
Make sure that you connect the Versa to a computer that meets or exceeds these specifications.
component
specification
Processor
300MHz Pentium II (Or Equivalent)
Memory
512MB
Hard Drive Space
Approximately 50MB
Display
VGA
Drive
4x CD-ROM
Operating System
Windows® XP Pro or greater
USB Port
Available
Introduction • Page 2-2
Always use instrument-grade Teflon® tape to seal thread
connections. Do not use pipe dope or lower grades of Teflon®
tape. Volatile materials in the dope and/or low grade tape
will contaminate the tubing!
Содержание 15-0800-074
Страница 1: ...Versa User Manual www teledynetekmar com Versa Manual Part 15 0800 074 ...
Страница 8: ...Versa User Manual www teledynetekmar com Chapter 1 Introduction ...
Страница 21: ...Versa User Manual www teledynetekmar com Chapter 2 Installation Setup ...
Страница 36: ...Versa User Manual www teledynetekmar com Chapter 3 Basic Operation ...
Страница 44: ...Versa User Manual www teledynetekmar com Chapter 4 Versa TekLink 2G ...
Страница 61: ...Versa User Manual www teledynetekmar com Chapter 5 21 CFR Part 11 Compliance ...
Страница 70: ...Versa User Manual www teledynetekmar com Installation Setup Page 2 6 Chapter 6 Method Development ...
Страница 77: ...Versa User Manual www teledynetekmar com Chapter 7 Maintenance Troubleshooting ...
Страница 90: ...Versa User Manual www teledynetekmar com Chapter 8 Diagrams ...
Страница 91: ...Versa User Manual www teledynetekmar com Detailed Plumbing Diagrams Page 8 2 ...
Страница 100: ...Versa User Manual www teledynetekmar com Chapter 9 Index ...