Installation
1- 6
VM700T Option 01 (NTSC) & Option 11 (PAL) User Manual
Benchtop Mounting.
For a standalone VM700T on a benchtop or for portable use,
apply the four adhesive rubber feet provided with the accessory to the bottom of
the box at the corners. Place the conditioner box in a convenient location where
you can make the power cord connections.
Power Cord Connections.
Connect the Power Line Conditioner in the power line
between the AC mains and the VM700T power input connector in the following
order:
1.
Plug the output power cord connected to the box into the AC power input
connector of the VM700T.
2.
Plug the AC mains power cord into the AC input connector of the power line
conditioner box.
3.
Plug the AC mains power cord into the 230 Vac mains source.
The measurement set, as received from the factory, is configured to skip
diagnostics at power on, and it starts up in the quick start mode with the
Waveform application running. When configured to skip the diagnostics, you can
run diagnostics at power on by pressing and holding in the Menu button while
turning on the front-panel power switch. The measurement set then starts up in
the low-level diagnostics mode with the VM700T startup menu as shown in
Figure 1--3. Use the following procedure to force an auto reset. All the diagnos-
tics then run at power on regardless of the power on diagnostic configuration.
1.
Turn off the measurement set.
2.
Hold down the
Menu
button and use the front panel STBY/ON switch to
power on the instrument.
3.
After the instrument beeps twice, release the
Menu
button. The VM700T
displays the VM700T Startup Menu.
4.
In the menu, use the control knob to select R ... INSTRUMENT RUN
MODES, and press the
Select Line
button. A second diagnostic menu is
then displayed.
5.
In this menu, select A ... AUTO RESET, and press the
Select Line
button.
This runs the full set of internal diagnostics. Check that all tests pass as they
run.
After the diagnostics complete, the measurement set initializes in the Waveform
mode unless there is a powerup function file (refer to Function Playback on
Power Up on page 1--3). If the any of the diagnostic tests fail, refer your
instrument to a qualified service person.
Power Up Diagnostics
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание VM700T
Страница 5: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 16: ...Getting Started Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 36: ...Operating Basics Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 74: ...Reference Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 75: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 106: ...Appendices Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 130: ...Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 131: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...