![Tektronix VM700T Скачать руководство пользователя страница 102](http://html1.mh-extra.com/html/tektronix/vm700t/vm700t_user-manual_1078247102.webp)
Auto Mode
VM700T Option 01 (NTSC) & Option 11 (PAL) User Manual
3- 27
The following aspects of Auto mode can be configured by the user:
H
Caution and Alarm Limits (Auto Limit Files)
H
locations of test signals on VITS lines and presence of zero carrier
(Measurement Locations file).
H
a subset of the available measurements (Selected Measurements file)
H
times for reports to be generated and the Selected Measurements file to be
used at each report time (Timed Events file)
Limit files for Auto mode are located in the Auto_Limit Files directory. This
directory is accessed by pressing the Configure button, selecting the Configure
Files soft key, and then selecting the Auto_Limit Files directory. There are five
limits file in the directory when the instrument is shipped from the factory.
Each limit file contains a list of all the measurements available for the Auto
mode. Each measurement can have limits that, when exceeded, produce caution
or alarm messages on the display. Both Caution and Alarm Limits can be set for
each measurement. The supplied limit files cannot be edited. Their contents can
be viewed by selecting (touching) a file. Only the Exit soft key is available to
close these files. Use the Create File soft key to copy the contents of one of the
supplied files to create a limit file you can customize.
The first line of the file is the title line. The second line of the file, “2 consecu-
tive error(s) before reporting,” sets the number of consecutive measurements that
find a value that is outside the Alarm Limits before logging the error to the
printer.
To change the consecutive error(s) before reporting value:
1.
Rotate the control knob to scroll the screen and highlight the 2 consecutive
error(s) before reporting line.
2.
Touch the number at the front of the line.
3.
Rotate the control knob to select either 1, 2, or 3.
4.
Select the Accept Input soft key or touch the screen inside the directory
window other than the selected parameter box to return the control knob to
its scrolling function.
To change a Caution or Alarm Lower or Upper Limit:
1.
Rotate the control knob to scroll the screen and highlight the line with the
limit you wish to change.
2.
Touch the limit you wish to change or remove.
Setting Alarm Limits
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание VM700T
Страница 5: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 16: ...Getting Started Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 36: ...Operating Basics Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 74: ...Reference Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 75: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 106: ...Appendices Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 130: ...Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 131: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...