Appendix B: Performance Verification
VM700T Option 01 (NTSC) & Option 11 (PAL) User Manual
B- 5
9.
Again, to speed up the response to the frequency change, toggle the Average
hard key off. Turn averaging on again after the display has re-stabilized.
10.
Check that the amplitude is within
±
0.55% of the reference level stored in
step 4.
11.
Change the TV Standard parameter from PAL to NTSC by pressing the XY
Arrow hard key. Change the filter selection from No Filter to the NTSC
bandwidth limiting filter by pressing the Move/Expand hard key until the
filter displays NTSC BW Lim.
12.
Return the leveled sine-wave generator to 50 kHz; do not change the output
level.
13.
To speed up the response to the frequency change, toggle the Average hard
key off. Turn averaging on again after the display has re-stabilized.
14.
Check that the amplitude is within
±
0.5% of the reference level stored in
step 4.
15.
Reset the leveled sine-wave generator for 3.6 MHz; do not change the output
level.
16.
To speed up the response to the frequency change, toggle the Average hard
key off. Turn averaging on again after the display has re-stabilized.
17.
Check that the amplitude is within
±
0.5% of the reference level stored in
step 4.
18.
Change the Filter parameter from NTSC BW Lim to Chroma BP by pressing
the Move/Expand hard key. Don’t change the settings on the leveled
sine-wave generator.
19.
To speed up the response to the filter change, toggle the Average hard key
off. Turn averaging on again after the display has re-stabilized.
20.
Check that the amplitude is within
±
0.5% of the reference level stored in
step 4.
21.
Change the TV standard parameter to PAL by pressing the XY Arrow hard
key.
22.
Change the leveled sine-wave generator frequency to 4.43 MHz; do not
change the output level.
23.
To speed up the response to the frequency change, toggle the Average hard
key off. Turn averaging on again after the display has re-stabilized.
24.
Check that the amplitude is within
±
0.5% of the reference level stored in
step 4.
25.
Repeats steps 1 through 7 for channel B and channel C.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание VM700T
Страница 5: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 16: ...Getting Started Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 36: ...Operating Basics Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 37: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 74: ...Reference Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 75: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 106: ...Appendices Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 130: ...Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 131: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...