5
Using TRIGGER Controls
TDS1000B and TDS2000B Series Oscilloscopes – Operator Training Kit
5-11
HF Reject
only allows signal components below 50 kHz to
pass to the trigger circuit. This helps to reduce the effects of
high frequency noise that might cause triggering problems.
Noise Reject
allows all components of a signal to pass to
the trigger circuit, but increases the peak-to-peak signal
required to trigger the signal. This mode reduces the
triggering sensitivity of the oscilloscope, and decreases the
chance of false triggering on noise.
LF Reject
is an AC-coupled triggering mode that passes
only the higher speed signal details above 50 kHz sine wave
bandwidth equivalent.
AC
trigger coupling blocks the DC component, such as a
bias voltage, and passes signals with a frequency greater
than 10 Hz.
You can now try the other coupling selections of the
oscilloscope and use the
TRIGGER VIEW
button to see the
effects of trigger coupling on the triggering signal.