Using VERTICAL Controls
3-24
TDS1000B and TDS2000B Series Oscilloscopes – Operator Training Kit
To perform an FFT operation by using the
MATH
MENU
button of the
VERTICAL
controls, follow these steps:
1. Connect
the
CH1
probe tip to the
5V@1kHz
pin and the
CH1
ground lead to the ground reference pin directly
below the
5V@1kHz
pin on the front panel.
2. Remove
the
CH2
probe tip from the Training 1 signal
board.
3. On the top of the front panel, push
DEFAULT SETUP
.
4. On the top of the front panel, push the
AUTOSET
button.
5. In
the
VERTICAL
section, push the
MATH MENU
button.
6. Push the appropriate side-screen menu button to select
Operation
FFT
and
Source CH1
.
7. In
the
HORIZONTAL
section, turn the
SEC/DIV
knob
counterclockwise to set the frequency scale at
5.00 kHz
.