5
Using TRIGGER Controls
TDS1000B and TDS2000B Series Oscilloscopes – Operator Training Kit
5-23
To perform the step response procedure, follow these
steps:
1. Connect the CH1 probe tip to the
FAST RISE TIME
signal on pin 16 and the CH1 ground lead to
GND
on
pin 1 of the Training 1 signal board.
2. On the top of the front panel, push the
DEFAULT
SETUP
menu button.
3. On the Training 1 signal board, push the
POWER
button
until the
Digital PWR
LED is on.
4. In
the
VERTICAL
section, turn the
CH1 VOLTS/DIV
knob clockwise for a readout of
200mV
.
5. In
the
VERTICAL
section, turn the
CH1
POSITION
knob
counterclockwise to position the trace two divisions
above the bottom of the display.
6. In the HORIZONTAL section, turn the
SEC/DIV
knob
clockwise for a time base readout of
M 5.00ns
.