This procedure is useful in scenarios where you need to
trouble shoot disc drives and remote controls. A typical
command signal of remote control has a 0.5 ms presignal
pulse followed by a delay of 3 ms and then followed by a 7
ms six-bit biphase code word. Setting the trigger pulse width
to 0.5 ms would reference the trigger point to the pre signal
pulse, thus ensuring a stable waveform display.
You can use the HELP menu, shown in another section of
this manual, to learn more about this special Pulse
triggering mode.