Installation
Quick Tip
To maintain measurement accuracy, degauss
your probe in each of these cases:
After you turn on the measurement system
and allow a 20-minute warm-up period
Before you connect the probe to a
conductor
Whenever a current or thermal overload
condition occurs
Whenever you subject the probe to a
strong external magnetic
fi
eld
4
TCP0030 Current Probe Instruction Manual