Performance Veri
fi
cation
1.
Connect the BNC cable to the output of the
pulse generator.
2.
Connect the other end of the BNC cable to
the HF current loop.
3.
Set the pulse generator output and pulse
width to maximum.
4.
Set the oscilloscope:
Vertical sensitivity to 200 mA/div
Horizontal to 2 ns/div
Trigger at 50%
Averaging on (32)
Coupling to DC
Automeasurement to Rise Time
5.
Set the probe range to the 5 A setting.
6.
Degauss the probe.
60
TCP0030 Current Probe Instruction Manual