Performance Veri
fi
cation
Measure the Oscilloscope Gain
1.
Connect a BNC-to-Dual Banana adapter
to the voltage output terminals of the
calibrator.
2.
Connect a BNC cable between the adapter
and the oscilloscope. (Leave the probe
connected to keep it warmed up.)
3.
Set the oscilloscope:
Vertical sensitivity to 500 mV/div
Vertical position and offset to 0.0 V
Horizontal to 400 or 500 us/div, trigger
50%
Acquisition mode to average 16
Input to 1 M
Ω
and coupling to DC
Automeasurement to display
AMPLITUDE