Maintenance
Reassembly
1.
Before reassembling the probe body, be sure to check:
a.
That the gap between the stationary and moveable core pieces is clean. If necessary, use isopropyl alcohol or a similar
cleaning agent to clean the pieces.
b.
That the contacts of the slide switch are clean, and if not, clean them.
c.
If the plastic slide assembly requires lubrication. If so, apply silicone-based grease sparingly to the parts.
2.
Reassemble the probe body by following steps 1 through 8 of the probe head disassembly procedure in reverse. (See
page 69,
NOTE.
Exercise care when
fi
tting the slide back into the probe body; aligning the switch contacts requires precision.
If the slider does not work smoothly after tightening the two screws, loosen the screws slightly.
76
TCP0030 Current Probe Instruction Manual