Maintenance
Do not use chemical cleaning agents; they may damage the probe. Avoid using chemicals that contain benzine, benzene,
toluene, xylene, acetone, or similar solvents.
Clean the exterior surfaces of the probe with a dry, lint-free cloth or a soft-bristle brush. If dirt remains, use a soft cloth or swab
dampened with a 75% isopropyl alcohol solution and rinse with deionized water. A swab is useful for cleaning narrow spaces on the
probe
,
use only enough solution to dampen the swab or cloth. Do not use abrasive compounds on any part of the probe.
Repairing the Probe
The subassemblies listed below can be replaced on the probe. The replacement procedures are on the following pages.
Replaceable component
Procedures to use
Transformer
Probe Head Disassembly, Transformer Replacement
Cable/Circuit Board Assembly
Probe Head Disassembly, Cable/Circuit Board Assembly Replacement,
Compensation Box Replacement
Compensation Box
Compensation Box Replacement
Switch Panel
Compensation Box Replacement
TCP0030 Current Probe Instruction Manual
67
Содержание TCP0030
Страница 1: ...TCP0030 120 MHz 30 A AC DC Current Probe Instruction Manual P071181200 071 1812 00 ...
Страница 2: ......
Страница 3: ...TCP0030 120 MHz 30 A AC DC Current Probe Instruction Manual www tektronix com 071 1812 00 ...
Страница 6: ......
Страница 10: ...Table of Contents iv TCP0030 Current Probe Instruction Manual ...
Страница 18: ...Preface xii TCP0030 Current Probe Instruction Manual ...
Страница 62: ...Specifications Figure 3 Maximum peak pulse versus pulse width 44 TCP0030 Current Probe Instruction Manual ...
Страница 64: ...Specifications 46 TCP0030 Current Probe Instruction Manual ...
Страница 97: ...Maintenance Figure 4 Replaceable Parts TCP0030 Current Probe Instruction Manual 79 ...