Performance Veri
fi
cation
16.
Set the calibrator to output a 500 Hz square
wave into the 5-turn current loop.
17.
Set the calibrator to the
fi
rst value in the
Test Record, and enable the output.
18.
Measure the waveform amplitude (c).
19.
Use the measured waveform amplitude (c)
from step 18 and the oscilloscope DC gain
accuracy value (b) from step 7 to calculate
the probe DC gain accuracy (d).
20.
Record the calculated probe DC gain
accuracy value (d) in the test record.
21.
Set the probe range to 30 A.
22.
Set the oscilloscope to 5 A/div.
23.
Measure the waveform amplitude (e).
5 A Range:
Example:
58
TCP0030 Current Probe Instruction Manual