Adjustments
Adjustments
The procedures that follow describe adjustments to the probe to bring the performance within the warranted speci
fi
cations.
DC gain accuracy
Equipment Required
Refer to the
Performance Veri
fi
cation
procedure for the equipment required. You also need an insulated,
fl
at-bladed adjustment tool.
Equipment Setup
Refer to the
Performance Veri
fi
cation
procedure for the equipment setup.
CAUTION.
To avoid ESD damage to the probe, use an antistatic wrist strap and work at a static-approved workstation when
handling the probe
TCP0030 Current Probe Instruction Manual
63
Содержание TCP0030
Страница 1: ...TCP0030 120 MHz 30 A AC DC Current Probe Instruction Manual P071181200 071 1812 00 ...
Страница 2: ......
Страница 3: ...TCP0030 120 MHz 30 A AC DC Current Probe Instruction Manual www tektronix com 071 1812 00 ...
Страница 6: ......
Страница 10: ...Table of Contents iv TCP0030 Current Probe Instruction Manual ...
Страница 18: ...Preface xii TCP0030 Current Probe Instruction Manual ...
Страница 62: ...Specifications Figure 3 Maximum peak pulse versus pulse width 44 TCP0030 Current Probe Instruction Manual ...
Страница 64: ...Specifications 46 TCP0030 Current Probe Instruction Manual ...
Страница 97: ...Maintenance Figure 4 Replaceable Parts TCP0030 Current Probe Instruction Manual 79 ...