Reference > Troubleshooting
Error and Information Messages
Disabled: Freq Mask Trigger in use
The DPX Spectrum display cannot run while the instrument is set to use Frequency Mask Trigger.
To run the DPX Spectrum display, you must turn off Frequency Mask Trigger.
DPX disabled: RefLev too low for Volts/Watts units
The Reference Level value is set too low. It is possible to set a RefLev value so low that DPX can’t run if
volts or watts are the selected amplitude units. Reset the Reference Level to a higher value.
Needs swept acq or larger AcqBW - Acquire data while display is selected
The display is not running because it needs to perform multiple acquisitions (it must be in swept mode) but
it is not the selected display. Only the selected display can perform multiple acquisitions.
Select the display showing this message to give it control of acquisitions parameters.
Sometimes, only one display can work and the others will be blank and show errors. This happens when
different displays have con
fl
icting demands on the acquisition data record.
This is a reminder that the
has been set to manual, which might affect the quality of
measurement results.
An error occurred while exporting Results. The
fi
le was not created.
Export the results again.
Export failure: unable to open results
le for export. File not saved.
The Export Results
fi
le could not be opened for writing, so the export of results was not completed.
Verify that there is suf
fi
cient free space on hard disk.
External Frequency Reference signal not valid.
The instrument is no longer locked to the external reference signal because it can no longer detect a signal
with proper amplitude and frequency.
Verify that the external reference signal is still connected to the instrument.
Hardware failure detected by Diagnostics
A hardware failure was detected while running Diagnostics. To see a description of the event initiated
by the failure,
display the Windows Event Viewer (see page 373)
. There is a shortcut to the Windows
Event Viewer on the Windows desktop.
Hardware failure - see Windows Event Viewer
A hardware failure has occurred. To see a description of the event initiated by the failure,
Windows Event Viewer (see page 373)
. There is an shortcut to the Windows Event Viewer on the
Windows desktop.
cient data for CISPR. Acquire while display is selected
RSA6100A Series Real-Time Spectrum Analyzers Help
369
Содержание RSA6100A Series
Страница 1: ...xx RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document P077016903 077 0169 03...
Страница 2: ......
Страница 3: ...RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document www tektronix com 077 0169 03...
Страница 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 285: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 273...
Страница 286: ...GP Digital Modulation Symbol Maps Symbol Maps 274 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 287: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 275...
Страница 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...
Страница 388: ...Reference Troubleshooting Displaying the Windows Event Viewer 376 RSA6100A Series Real Time Spectrum Analyzers Help...