![Tektronix RSA6100A Series Скачать руководство пользователя страница 337](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169337.webp)
Controlling the Acquisition of Data > Using Triggers to
Capture Just What You Want > Triggering
Event Tab
Source
Source speci
fi
es the source of the signal to be monitored for a trigger event. The choices available are:
RF Input, Trig In (front), Trig 2 In (rear), Gated, and Line. The Source drop-down list is always visible
on the Event tab.
Type
Type de
fi
nes the trigger type as Power , Frequency Mask, DPX Density, or Runt. The Type selection
is only present when Source is set to RF Input.
When Type is set to Power, a transition through an adjustable level of the selected source determines
a trigger event.
When Type is set to Frequency Mask, trigger events are de
fi
ned by where and how the input signal
crosses or fails to cross the edge of the mask.
When Type is set to DPX Density, the amount of time a signal is present at a particular point in the
DPX Spectrum graph de
fi
nes a trigger event.
When Type is set to Runt, a signal that exceeds a speci
fi
ed level and then fails to exceed a second
speci
fi
ed level de
fi
nes a trigger event.
Source = RF input.
The following table describes the settings for each triggering type available when Source is set to RF Input.
Type speci
fi
es whether a Power level, Frequency Mask violation, DPX Density threshold violation, or
Runt signal will de
fi
ne a trigger event.
Table 5: Trigger types
Event Type
Setting
Description
Power
For the Power setting, if the RF input signal crosses the speci
fi
ed
level (in the direction set by Slope), a trigger event occurs. A trigger
event does not occur if the power level of the signal simply exceeds
the level setting, there must be a power level transient that crosses
the speci
fi
ed level.
Slope
The slope control determines whether the spectrum analyzer
fi
nds
the trigger point on the rising or falling edge of the RF input signal.
Level
The Level setting determines the power level that will be recognized
as a trigger event. The Level is indicated in the Amplitude vs Time
and Time Overview displays with a dark cyan colored-line.
RSA6100A Series Real-Time Spectrum Analyzers Help
325
Содержание RSA6100A Series
Страница 1: ...xx RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document P077016903 077 0169 03...
Страница 2: ......
Страница 3: ...RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document www tektronix com 077 0169 03...
Страница 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 285: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 273...
Страница 286: ...GP Digital Modulation Symbol Maps Symbol Maps 274 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 287: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 275...
Страница 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...
Страница 388: ...Reference Troubleshooting Displaying the Windows Event Viewer 376 RSA6100A Series Real Time Spectrum Analyzers Help...