Analog Modulation > PM
PM Display
Item
Element
Description
1
Check mark indicator
The check mark indicator in the upper, left-hand corner of the display shows
when the PM display is the optimized display.
2
Top of Graph control
Sets the phase value indicated at the top of the graph. Since the Position
value at the vertical center of this graph remains constant as the Top of Graph
value is adjusted, the Vertical Scale increases as the Top of Graph value
increases, which also affects the bottom of graph readout. Vertical Scale can
also be controlled from the Settings control panel’s Scale tab.
3
Position
Speci
fi
es the phase shown at the center of the graph display. Changing this
value moves the trace up and down in the graph, which affects the Top of
Graph and Bottom of Graph values as well.
4
Bottom of Graph Readout
Displays the phase value at the bottom of the graph.
5
Measurement readouts
Displays numeric values for the +Pk (positive phase deviation), –Pk (negative
phase deviation), Pk-Pk (peak-to-peak phase deviation), and RMS (RMS
value of the phase deviation).
6
Position
Speci
fi
es the horizontal position of the trace on the graph display.
7
Scale
Adjusts the time range of the graph. By decreasing the scale (full-scale time
over 10 divisions), the graph essentially becomes a window that you can
move over the acquisition record by adjusting the horizontal position.
8
Clear
Clears the measurement results.
9
Trace Detection readout
Displays the Trace Detection setting (see Settings > Trace > Detection).
10
Freq Error
This readout can show Freq Error or Freq Offset. When it displays Freq
Error, it shows the difference between the instrument Frequency setting and
the measured value of the signal’s carrier frequency. When it displays Freq
Offset, it shows the frequency offset speci
fi
ed on the Settings > Parameters
tab. If Freq Error is displayed, it also indicates that the Carrier frequency
detection setting is Auto. If Freq Offset is displayed, it indicates that the
Carrier frequency detection setting is manual.
xxx
Changing the PM Settings (see page 126)
RSA6100A Series Real-Time Spectrum Analyzers Help
125
Содержание RSA6100A Series
Страница 1: ...xx RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document P077016903 077 0169 03...
Страница 2: ......
Страница 3: ...RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document www tektronix com 077 0169 03...
Страница 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 285: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 273...
Страница 286: ...GP Digital Modulation Symbol Maps Symbol Maps 274 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 287: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 275...
Страница 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...
Страница 388: ...Reference Troubleshooting Displaying the Windows Event Viewer 376 RSA6100A Series Real Time Spectrum Analyzers Help...