![Tektronix RSA6100A Series Скачать руководство пользователя страница 155](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169155.webp)
RF Measurements > MCPR
Freq & RBW Tab for ACPR and MCPR Displays
Freq & RBW Tab for ACPR and MCPR Displays
The Freq & RBW tab speci
fi
es frequency parameters for the Channel Power & ACPR measurements and
MCPR measurement.
Setting
Description
Center Freq
Speci
fi
es the center/measurement frequency.
Step
The Step control sets the increment/decrement size for the knob adjustment of the
center frequency. Arrow key adjustment of frequency is also affected (it is 10x the knob
adjustment). If Auto is enabled, the spectrum analyzer will adjust the Step size as
required.
RBW
Select Auto or Manual. Adjusts the RBW for the entire measurement. This setting is
Independent of the Spectrum view’s RBW setting.
VBW
Adjusts the VBW (Video Bandwidth) value. VBW Maximum = current RBW value. VBW
Minimum = 1/10,000 RBW setting.
xxx
NOTE.
While the center frequency and step settings on this control tab are shared between ACPR, MCPR
and Occupied Bandwidth, the RBW setting is not shared. The RBW setting is unique for each measurement.
VBW
The VBW setting enables/disables the Video Bandwidth
fi
lter. VBW is used in traditional swept analyzers
to reduce the effect of noise on the displayed signal. The VBW algorithm in the analyzer emulates the
VBW
fi
lters of traditional swept analyzers.
The maximum VBW value is the current RBW setting. The minimum VBW value is 1/10,000 of the
RBW setting.
NOTE.
If you are following a procedure that says to “set VBW to three times the RBW value or greater”, it
means that the test should be conducted with no VBW effects. In the RSA6100A Series instruments, this
condition is met by disabling the VBW function.
RSA6100A Series Real-Time Spectrum Analyzers Help
143
Содержание RSA6100A Series
Страница 1: ...xx RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document P077016903 077 0169 03...
Страница 2: ......
Страница 3: ...RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document www tektronix com 077 0169 03...
Страница 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 285: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 273...
Страница 286: ...GP Digital Modulation Symbol Maps Symbol Maps 274 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 287: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 275...
Страница 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...
Страница 388: ...Reference Troubleshooting Displaying the Windows Event Viewer 376 RSA6100A Series Real Time Spectrum Analyzers Help...