Reference > Troubleshooting
Error and Information Messages
This message appears when a measurement is not the selected measurement and CISPR
fi
lters are selected.
Set the measurement to be the selected measurement and reacquire the signal.
This message occurs in GP Digital Modulation displays. The most likely cause for this message is that
there are not enough symbols to analyze. This can happen if
The Analysis Length is set too short. Increase the Analysis Length on the Analysis Time tab.
The Analysis Offset has pushed the Analysis Time so far out that the actual Analysis Length is too
short, even though the user-requested Analysis Length would have been long enough without the
excessive offset. Decrease the Analysis Offset on the Analysis Time tab.
The input signal is bursted, and the burst does not contain enough symbols.
Locking to ext freq ref signal...
This a status message displayed while the instrument is attempting to lock to an external reference signal.
The message disappears when the instrument locks to the reference signal.
Needs swept acq or larger Acq BW - Acquire data while display is selected
The display has one of two problems: It is not the selected display, which prevents it from controlling
the hardware acquisition parameters, and setting the acquisition mode to Swept; or its settings require a
wider data bandwidth.
Select the display that you are interested in and it will change the acquisition to meet its own needs.
Increase the acquisition bandwidth manually or by changing the selected display’s settings to cause
the wider bandwidth.
No Math trace: unmatched trace lengths
A math trace could not produced because the traces selected to generate the math trace do not have
the same number of points. This can easily happen if both traces are recalled, but were saved under
different “Points” settings. This can also occur if one of the selected traces is a live trace and the other
trace is a recalled trace.
In a Spectrum display, as long as one trace is live, you can change the “Points” setting (Setup >
Settings > Freq & Span tab) to match the recalled trace.
If you are using two recalled traces to generate the Math trace. You must recreate at least one of
the traces.
The instrument needs to be aligned. When the instrument is not aligned, measurements might not be
accurate.
Select Tools > Alignments > Align Now to alignment the instrument.
The
Burst Detection Mode (see page 265)
is On, but no burst was detected in the signal.
Check that the Threshold setting is properly set.
370
RSA6100A Series Real-Time Spectrum Analyzers Help
Содержание RSA6100A Series
Страница 1: ...xx RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document P077016903 077 0169 03...
Страница 2: ......
Страница 3: ...RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document www tektronix com 077 0169 03...
Страница 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 285: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 273...
Страница 286: ...GP Digital Modulation Symbol Maps Symbol Maps 274 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 287: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 275...
Страница 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...
Страница 388: ...Reference Troubleshooting Displaying the Windows Event Viewer 376 RSA6100A Series Real Time Spectrum Analyzers Help...