Performance Verification
4.
Set the RSA5100B:
Center Frequency
Setup > Settings > Freq & Span > Center
1 GHz
Span
Setup > Settings > Freq & Span > Span
25 MHz
RBW
Setup > Settings > BW > RBW
1 kHz
Function
Setup > Settings > Traces > Function
Avg (RMS)
Averages
Setup > Settings > Traces > Function
25 (count box is checked)
Ref Level
Setup > Amplitude > Ref Level
-15 dBm
Internal Attenuator
Setup > Amplitude > Internal Attenuator
10 dB (Auto unchecked)
5.
Set the RF generator:
Output Level
–15 dBm
Output Frequency
1 GHz
RF
On
6.
Turn on the Marker Reference (MR) and Marker 1 (M1) and set them for
Delta Operation.
Select
Markers
located at the bottom right of the screen on the status bar
to display the Markers control bar.
Click the down arrow on the left side of the Markers control bar to view
the Select a Marker dropdown menu. Select
Add Marker
to add the
MR marker.
Select
Add Marker
again to add the M1 marker.
Select
Define
located on the right side of the Markers control bar to open
the Define Markers control panel.
Select
Delta
from the
Readouts
dropdown menu.
7.
After averaging has completed, select Reference Marker
MR
using the Select
a Marker dropdown menu on the left side of the Markers control bar.
8.
Press the
Peak
key on the Markers control bar.
9.
Select Marker 1
M1
using the Select a Marker dropdown menu on the left
side of the Markers control bar.
RSA5100B Series Technical Reference