Performance Verification
DANL - Preamp ON, LF
Path (Option 50/51 Only)
1.
Reset the RSA5100B to factory defaults: select
Setup
>
Preset (Main)
.
2.
Select
Tools
>
Alignments
and select
Align Now
.
3.
Set the RSA5100B:
Reference Level
Setup > Amplitude > Internal Settings > Ref Level
–50 dBm
Internal Attenuator
Setup > Amplitude > Internal Settings > Internal
Attenuator
0 dB (Auto unchecked)
RF & IF Optimization
Setup > Amplitude > RF & IF Optimization
Minimize Noise
Internal Preamp
Setup > Amplitude > Internal Settings > Internal
Preamp
ON (Internal Preamp box checked)
Center Frequency
Setup > Settings > Freq & Span > Center
1.1 MHz
Span
Setup > Settings > Freq & Span > span
100 kHz
RBW
Setup > Settings > BW > RBW
Auto (box checked)
Detection
Setup > Settings > Traces > Detection
Avg (of logs)
Function
Setup > Settings > Traces > Function
Avg (of logs)
Count
Setup > Settings > Traces > Function
100 (Count box checked)
LF Path
Setup > Acquire > Input Params
Use Low Freq... (box checked)
4.
Set the markers for Noise Mode (dBm/Hz) operation:
a.
Select
Markers
located at the bottom right of the screen on the status bar
to display the Markers control bar.
b.
Click the down arrow on the left side of the Markers control bar to view
the Select a Marker dropdown menu. Select
Add Marker
to add the
MR marker.
c.
Select
Add Marker
again to add the M1 marker.
RSA5100B Series Technical Reference