Performance Verification
Table 78: Frequencies of interest for DANL check (Option 50/51)
Center frequency
Marker Noise level
Frequency range
1.1 MHz
9.9 MHz
1 MHz - 10 MHz (All models)
10.1 MHz
1.99 GHz
10 MHz - 2 GHz (All models)
2.01 GHz
2.99 GHz
2 GHz - 3 GHz (All models)
3.01 GHz
6.2 GHz
3 GHz - 6.2 GHz
(RSA5106B/5115B/5126B)
6.2 GHz
15 GHz
6.2 GHz - 15 GHz
(RSA5115B/5126B)
15 GHz
26.5 GHz
15 GHz - 26.5 GHz
(RSA5115B/5126B)
7.
Enter the highest noise level for each of the frequency ranges shown into the
test record. (Limits are shown in the test record.)
IF Flatness (Channel Response)
1.
Connect the RF generator, power splitter, power meter, and RSA5100B as
shown in the following figure.
The power splitter outputs should connect directly to the RSA5100B RF Input
and to the Power Sensor.
Figure 14: Equipment connections for IF
fl
atness check
2.
Reset the RSA5100B to factory defaults: select
Setup
>
Preset (Main)
.
120
RSA5100B Series Technical Reference