Specifications
3
Option B16x, B16xHD only.
4
Option B125, B125HD only.
Table 16: Channel response after cal/norm (center frequency 65 MHz) (RSA5115B/5126B)
Characteristic
Description
Amplitude Flatness
Speci
fi
cation
Span
≤
300 kHz
±0.5 dB
1
1
Checked by the IF test at 65 MHz combined with the RF
fl
atness.
Table 17: Pulse measurements, typical
Characteristic
Description
85, 125, and 160 MHz BW
25 and 40 MHz BW
Minimum Pulse Width for detection, typical
50 ns
150 ns
±0.3 dB + absolute Amplitude Accuracy
Average ON Power
(18 °C to 28 °C), typical
For pulse widths
≥
100 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
For pulse widths
≥
300 ns, and signal
levels >70 dB below Ref Level
±0.2% of reading
Duty Factor, typical
For pulse widths
≥
150 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
For pulse widths
≥
450 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
±0.4 dB + absolute Amplitude Accuracy
Average Transmitted Power, typical
For pulse widths
≥
100 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
For pulse widths
≥
300 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
±0.4 dB + absolute Amplitude Accuracy
Peak Pulse Power, typical
For pulse widths
≥
100 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
For pulse widths
≥
300 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
±0.25% of reading
Pulse Width, typical
For pulse widths
≥
150 ns, duty cycles
of 0.5 to 0.001, and signal levels >50 dB
below Ref Level
For pulse widths
≥
450 ns, duty cycles
of 0.5 to 0.001, and S/N ratio
≥
30 dB
System Rise time, typical
<12 ns (85, 125 MHz BW)
<7 ns (160 MHz BW)
<40 ns (25 MHz BW)
<25 ns (40 MHz BW)
RSA5100B Series Technical Reference