Performance Verification
Spurious Response
Residual Response
1.
Terminate the RSA5100B RF Input.
2.
Reset the RSA5100B to factory defaults: select
Setup
>
Preset (Main)
.
3.
Select
Tools
>
Alignments
and select
Align Now
.
4.
Set the RSA5100B:
Ref Level
Setup > Amplitude >
-40 dBm
Internal Attenuator
Setup > amplitude > Internal Settings > Internal
Attenuator
0 dB (Auto unchecked)
Setup > Amplitude > RF & IF Optimization >
Maximum Dynamic Range
RBW
Setup > Settings > BW > RBW
1 kHz
Function
Setup > Settings > Traces > Function
Avg (VRMS)
Count
Setup > Settings > Traces > Function > Avg (of
logs)
10 (Count checked)
5.
Set the Frequency (Freq key) to each center frequency in the Residual
Response Center Frequencies table. Set the span for each center frequency
as shown in the table. (See Table 80.) Wait for the averaging to complete,
then press the
Peak
key on the Markers control bar and record the marker
amplitude in the table.
6.
Enter the highest of these signal levels into the test record.
Table 80: Residual response center frequencies
Center
frequency
Span
Range
Marker amplitude
All models
600 MHz
800 MHz
200 MHz to 1 GHz
1.5 GHz
1 GHz
1 GHz to 2 GHz
2.5 GHz
1 GHz
2 GHz to 3 GHz
RSA5106B/5115B/5126B only
3.5 GHz
1 GHz
3 GHz to 4 GHz
4.5 GHz
1 GHz
4 GHz to 5 GHz
5.6 GHz
1.2 GHz
5 GHz to 6.2 GHz
RSA5115B/5126B only