Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-145
scan.mode
This attribute controls the scan mode setting.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
System reset
Scan reset
Create configuration script
0 (scan.MODE_OPEN_ALL)
Usage
scanModeSetting
= scan.mode
scan.mode =
scanModeSetting
scanModeSetting
The present scan mode setting. Set to one of the following values:
•
scan.MODE_OPEN_ALL
or
0
•
scan.MODE_OPEN_SELECTIVE
or
1
: See Details
•
Details
When this attribute is set to
scan.MODE_OPEN_ALL
, all channels on all slots are opened before a
scan starts.
When this attribute is set to
scan.MODE_OPEN_SELECTIVE
, an intelligent open is performed:
•
All channels involved in scanning are opened
•
Closed channels not involved in scanning remain closed during the scan
Example
scan.mode = scan.MODE_OPEN_SELECTIVE
Sets the scan mode setting to open
selective.
Also see
(on page 7-146)
(on page 3-1)