Section 7: TSP command reference
Models 707B and 708B Switching Matrix Reference Manual
7-154
707B-901-01 Rev. B / January 2015
trigger.blender[N].EVENT_ID or 58
to 59
A combination of events has occurred.
trigger.timer[N].EVENT_ID or 20
to 23
A delay expired.
tsplink.trigger[N].EVENT_ID or 17
to 19
An edge (either rising, falling, or either based on the
configuration of the line) on the TSP-Link trigger line.
lan.trigger[N].EVENT_ID or 29 to
36
A LAN trigger event has occurred.
scan.trigger.EVENT_SCAN_READY or
24
Scan ready event.
scan.trigger.EVENT_SCAN_START or
25
Scan start event.
scan.trigger.EVENT_CHANNEL_READY
or 28
Channel ready event.
scan.trigger.EVENT_MEASURE_COMP
or 56
Measure complete event.
scan.trigger.EVENT_SEQUENCE_COMP
or 50
Sequence complete event.
scan.trigger.EVENT_SCAN_COMP or
26
Scan complete event.
scan.trigger.EVENT_IDLE or 27
Idle event.
schedule.alarm[N].EVENT_ID or 37
to 38
Trigger event generated by the alarm
N
.
Use one of the text trigger event IDs (for example,
digio.trigger[
N
].EVENT_ID)
to set the
stimulus value rather than the numeric value. Doing this will make the code compatible for future
upgrades.
Example 1
scan.trigger.channel.stimulus =
scan.trigger.EVENT_SCAN_START
Sets the trigger stimulus of
the channel event detector
to scan start event.
Example 2
scan.trigger.channel.stimulus = 0
print(scan.trigger.channel.stimulus)
Starts the channel action
immediately after the Scan
Start Event. This also
resets the stimulus to the
default.
Output:
5.00001
Also see
(on page 7-152)
(on page 7-152)
(on page 3-1)
(on page 3-1)