Section 5: Configure and use a Series 700 Switching System
Model 4200A-SCS Parameter Analyzer User's Manual
5-14
4200A-900-01 Rev. C / January 2019
In this example, assigning TermIdStr1 to
SMU1
and Pin 1 to
3
connects SMU1 to Pin 3 on the matrix.
Figure 66: connectpins device connections
If the OpenAll check box is selected, the
connectpins
action opens all crosspoints before closing
the specified pairs. This is the default and is usually the preferred behavior. However, since
connectpins
only has eight field pairs, the action can only close eight crosspoints during each run.
To close more crosspoints, use multiple
connectpins