Section 8: Use the 4200A-CVIV Multi-Switch
Model 4200A-SCS Parameter Analyzer User's Manual
8-12
4200A-900-01 Rev. C / January 2019
2. Select
Tests
.
3. To find the test, type
vds-id
into the search bar, then select
Search
.
4. Select the
vds-id
test, then select
Add
to copy it to the project tree.
5. Select
Configure
.
Figure 115: Configure highlighted
6. Configure the source and measure parameters of SMU 1 (Source), SMU 2 (Drain), and SMU 3
(Gate) as needed.
7. Assign the instrument of the Bulk terminal to be
SMU 4
and set its Operation Mode to
Voltage
Bias
.
Figure 116: Bulk channel 4 SMU voltage bias configuration
8. Verify that the Bias is set to
0 V
and the Compliance is set to
0.1 A
.
Add an action for switching the CVU to the output terminals of
the 4200A-CVIV
To add an action:
1. Choose
Select
to return to the Library.
Figure 117: Select highlighted