![Tektronix KEITHLEY 4200A Скачать руководство пользователя страница 122](http://html1.mh-extra.com/html/tektronix/keithley-4200a/keithley-4200a_user-manual_1078018122.webp)
Model 4200A-SCS Parameter Analyzer User's Manual
Section 9: PMU for pulsed I-V measurements on a MOSFET
4200A-900-01 Rev. C / January 2019
9-7
Configure the test
To configure the test:
1.
Select
Configure
.
Figure 131: Configure button highlighted
2.
From the Key Parameters pane, adjust the pulse voltage sweep levels of PMU1-2 on the Drain
terminal and the Pulse Step Voltage of PMU1-1 on the Gate terminal, as needed.
Figure 132: Key Parameters pane for the pulse-vds-id test