![Tektronix KEITHLEY 4200A Скачать руководство пользователя страница 115](http://html1.mh-extra.com/html/tektronix/keithley-4200a/keithley-4200a_user-manual_1078018115.webp)
Model 4200A-SCS Parameter Analyzer User's Manual
Section 8: Use the 4200A-CVIV Multi-Switch
4200A-900-01 Rev. C / January 2019
8-17
Run the project and review the results
To run the project and review the results:
1.
Select the test name at the top of the project tree. In this example, it is
CVIV_COMPENSATION
.
2.
Select
Run
to start the test. The two actions and two tests run sequentially, and the 4200A-CVIV
is configured by the
cviv-configure
action.
3.
Select
Analyze
to view the results.
Figure 125: Analyze highlighted
4.
Select the
cv-nmosfet
test to view the results of the C-V test.
5.
Select the
vds-id
test to view the results of the I-V test.