Section 2: Getting started
Model 4200A-SCS Parameter Analyzer User's Manual
2-10
4200A-900-01 Rev. C / January 2019
Figure 7: Interlock connector on the rear panel of the 4200A-SCS
2. Connect the other end of the interlock cable to a compatible test fixture, such as the Keithley
Instruments LR:8028.
For more information, see the "Configuring the safety interlock" topic in the
4200A-SCS Reference
Manual
.
SMU connections
The following topics explain how to connect the source-measure units (SMUs) to the device under
test (DUT).
Do not touch test cables or connectors when powering up the 4200A-SCS. Hazardous voltage
may be output momentarily, posing a safety hazard that could result in personal injury or
death.
Do not turn on the 4200A-SCS until you have reviewed the safe power-up procedure in
(on page 2-15).
Do not connect the DUT to the 4200A-SCS before powering it up, because the hazardous
voltage that may be output momentarily at power-up could damage the DUT.
If your 4200A-SCS includes preamplifiers, all tests should be performed using the preamplifiers, as
the installed SMUs were optimized at the factory to use them.