Section 3: Create a new project and test
Model 4200A-SCS Parameter Analyzer User's Manual
3-8
4200A-900-01 Rev. C / January 2019
3.
Set the Gate terminal connection to
SMU2
.
4.
Set the Operation Mode to
Voltage Step
.
5.
Change the Start, Stop, Step, and Compliance settings to match the next figure or to the gate
settings appropriate for your device.
Figure 23: SMU2 steps from 2 V to 5 V, connected to MOSFET Gate terminal
6.
Set the Drain terminal connection to
SMU1
.
7.
Set the Operation Mode to
Voltage Linear Sweep
.
8.
Change the Start, Stop, Step, and Compliance settings to match the next figure.
Figure 24: SMU1 sweeps from 0 V to 5 V, connected to MOSFET Drain terminal
9.
Set the Operation Mode of the Source terminal to
GNDU
.