Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
3. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls:
a. Push UTILITY (front-panel)
System (bottom)
Factory Reset
(side)
OK (side).
b. Load the MODE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
c.
Push HORIZONTAL MENU (front-panel)
Clock (side).
d. Push 2, 0, 0 and M (SHIFT+7) keys in this order or turn the general pur-
pose knob to set the internal clock frequency to 200 MHz.
4. Push the RUN button.
The LED above the RUN button is on.
5. Verify that the clock signal amplitude is equal to or larger than 0.5 V, and the
clock signal period is 20 ns (50 MHz).
6. Disconnect the cable from the 1/4 Clock connector.
Содержание AWG610
Страница 4: ......
Страница 24: ...Preface AWG610 Arbitrary Waveform Generator User Manual...
Страница 94: ...Tutorials 2 56 AWG610 Arbitrary Waveform Generator User Manual Figure 2 36 The Standard Function dialog box...
Страница 216: ...Quick Editing AWG610 Arbitrary Waveform Generator User Manual...
Страница 220: ...The Table Editor AWG610 Arbitrary Waveform Generator User Manual...
Страница 242: ...The Sequence Editor AWG610 Arbitrary Waveform Generator User Manual...
Страница 288: ...The UTILITY Window AWG610 Arbitrary Waveform Generator User Manual...
Страница 292: ...Capturing Waveforms AWG610 Arbitrary Waveform Generator User Manual...
Страница 325: ...Programming Examples AWG610 Arbitrary Waveform Generator User Manual 3 207 Figure 3 66 Gaussian noise and ramp waveforms...
Страница 326: ...Programming Examples AWG610 Arbitrary Waveform Generator User Manual...
Страница 338: ...File Management AWG610 Arbitrary Waveform Generator User Manual...
Страница 424: ...Appendix B Performance Verification AWG610 Arbitrary Waveform Generator User Manual...
Страница 440: ...Appendix E File Transfer Interface Outline AWG610 Arbitrary Waveform Generator User Manual...