Theory of Operation—2246 1Y and 2246 Mod A Service
cannot move the voltage on U1102B pin 5 across the
comparison threshold (ground on pin 6).
During the vertical field sync pulse, the frequency of
the serration pulses (line and equalizing) doubles.
The filter capacitors will then be discharged enough
to go below ground and switch the output state of
U1102B. That signal is applied to U1104A pin 1 to be
available as the TV FIELD Sync trigger signal for the
A Trigger system.
The readout interface circuit responds to the
readout request and readout blanking inputs, and
generates a blanking signal (BLANK, pin 18) to
control the Z-Axis Amplifier enabling signals from
U602. The chop blanking signal also gets routed
through this circuit.
Pin Description
SYNC SWITCHING. Solid-state switches U1104A
and U1106C provide switching between the TV FIELD
and the TV LINE signal for the A Trigger and between
TV LINE from A SOURCE and the average DC level of
the measurement channel for the B Trigger. The
switching states are controlled by the Measurement
Processor via the TV FIELD SEL and the B TV TRIG
EN control signals from the Auxiliary Control Register
(U1103).
DISPLAY AND TRIGGER LOGIC AND
PROCESSOR INTERFACE (Diagram 4)
The Display Sequencer or SLIC (slow-logic inte
grated circuit, U600) performs most of the slow
logic functions required to run the display functions.
This integrated circuit contains a microprocessor
interface, the display sequencer logic circuitry, the
trigger holdoff tim er, the chop clock, and an inter
face to the on-screen readout control logic.
The microprocessor interface of U600 provides the
capability to serially load the internal control
register, write the internal read/write memory, do
some limited real-tim e control over a few sequencer
functions, and monitor status information.
The Display Sequencer contains a read/write
memory for storing the display states to be
sequenced through and logic for sequencing the A
and B Sweep displays and trigger sources. The
sequencer also provides control signals that are
needed to do waveform measurements.
An internal trigger holdoff tim er provides a pulse with
programmable width that is triggered on at the end
of A Sweep (or at the end of B Sweep). The pulse
width may be set from 1
jj
.
s
to greater than 0.5 s,
depending on the internal counter divide ratio, and
the holdoff oscillator frequency at pin 15.
The chop clock circuit generates a phase-dithered
chop clock and blanking signal, derived from an
external frequency source. With 10 MHz applied, the
chop rate can be 1.25 MHz or 625 kHz, with a
blanking tim e of about 200 ns (625 kHz is used in the
2246 1Y and 2246 Mod A).
The following is a description of Display Sequencer
U600 pin functions
(see Figure 3-1
for pin
num bers).
i
2
3
4
5
6
7
8
g
10
l i
12
13
14
15
16
17
18
19
20
(
6081
-
04
)
6555-31
TC
V DO
LFC
CH 1 EN
WR
CH 2 EN
A0
CH 3 EN
A1
D
CH 4 EN
A2
I
s
MGE
A3
P
ZEN
L
RD
A
ATS 2
DIO
Y
ATS 1
TDI
S
ATS 0
R0R
t
A SLOPE
Q
ROB
U
BTS 2
B GATE
E
BTS 1
N
A GATE
C
BTS 0
0SC0UT
E
R
B SLOPE
0SCRST
(SLIC)
OS
TH0
HD0
BLANK
HD1
S0UT
TS
vss
TEST
40
39
38
37
36
35
34
33
32
31
30
29
28
27
26
25
24
23
22
21
3-16
Figure 3-1. Display Sequencer 1C (SLIC, U600)
p in -o u t diagram.
Содержание 2246 1Y
Страница 13: ...2246A Service 7062 01 X The 2246 1Y or 2246 Mod A Portable Oscilloscope ...
Страница 35: ......
Страница 69: ...Theory of Operation 2246 1Y and 2246 Mod A Service 15 V 6081 07 3 32 Figure 3 4 Simplified Sweep Circuit ...
Страница 91: ...Theory of Operation 2246 1Y and 2246 Mod A Service 6081 12 6557 99 3 54 Figure 3 9 Power Supply block diagram ...
Страница 139: ......
Страница 185: ......
Страница 187: ......
Страница 278: ...Figure 9 5b Detailed 2246 1Y or 2246 Mod A block diagram part 2 R E ...
Страница 287: ......
Страница 304: ...A D D J U N 1991 ...
Страница 311: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 1 0 V r l III B f e I l i 6555 39 ...
Страница 314: ...15 14 13 12 11 10 9 8 7 6 5 12 11 10 9 8 7 AT117 A TI 271 1 2 3 4 5 6 6555 70 Figure 9 8 Hybrid pin identifiers ...
Страница 319: ......
Страница 320: ......
Страница 321: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 2 SET READOUT CONTROL CCW OFF 6555 40 ...
Страница 323: ......
Страница 324: ......
Страница 326: ...A B C P E F G ...
Страница 327: ......
Страница 331: ......
Страница 337: ......
Страница 340: ......
Страница 343: ...C l rti ...
Страница 344: ...WAVEFORMS FOR DIAGRAM 5 2246 1Y and 2246 Mod A Service HORIZ MODE A A SEC DIV 2M S HOLDOFF MIN CCW ...
Страница 345: ...HORIZ MODE ALT A SEC DIV 2M s B SEC DIV 5 mS HOLDOFF MIN DELAY INTENSIFIED ZONE STARTS MIDSCREEN ...
Страница 347: ...BEV JUNE 1991 7062 18 ...
Страница 349: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 6 6555 43 I ...
Страница 354: ...A L B i C 1 D 1 E i F i G 1 2 3 4 5 6 7 8 2246 1Y AND 2246 MOD A _________A fc A A REV JUNE 1991 7062 19 ...
Страница 355: ...I BOARD LOCATION 8K 8J 8H 7H 8H 8H 7J 8D 7H 8H 8J 8J 7K 8K 10D 6L 6K 6L 6K 6G 5H I ...
Страница 362: ......
Страница 372: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 9 MORE ...
Страница 373: ... 6555 72 6557 88 ...
Страница 375: ......
Страница 376: ...SCHEM LOCATION BOARD LOCATION 2E 2B 7F 4B 4J 5D 3J 5C 4L 6D 3L 6C 5L 7D 3L 7D 3M 7D 5M 7D IB 1A 1C 1A 2B 1A 3B 1A ...
Страница 380: ...WAVEFORMS FOR DIAGRAM 10 ...
Страница 387: ...WAVEFORMS FOR DIAGRAM 11 6555 73 6557 48 ...
Страница 394: ...WAVEFORMS FOR DIAGRAM 12 2246 1Y and 2246 Mod A Service i f I S 1 f c i 4 i ...
Страница 395: ...WAVEFORMS FOR DIAGRAM 12 cont 2246 1Y and 2246 Mod A Service ...
Страница 397: ......
Страница 398: ......
Страница 400: ...REV JUNE 1991 7062 39 ...
Страница 401: ......
Страница 403: ......
Страница 404: ......
Страница 406: ...B 2246 IV AND 2246 MOD A ...
Страница 407: ......
Страница 408: ......
Страница 414: ... _ 5 ...
Страница 415: ...2246 1Y 2246 MOD A ...
Страница 416: ......
Страница 423: ...2246 1Y 2246 MOD A ...
Страница 424: ...2246 1Y 2246 MOD A ...
Страница 425: ......
Страница 433: ......