Section 4—2246 1Y and 2246 Mod A Service
PERFORMANCE CHECK PROCEDURE
INTRODUCTION
PURPOSE
This Performance Check Procedure verifies the Perform
ance Requirements of the 2246 1Y or 2246 Mod A as
listed in the Specification (Section 1) and helps deter
mine the need for readjustment. These checks may also
be used as an acceptance test or as a preliminary
troubleshooting aid.
You do not have to remove the wrap-around cabinet from
the 2246 1Y or 2246 Mod A to do this procedure. All
checks can be made with controls and connectors ac
cessible from the outside.
TEST EQUIPMENT REQUIRED
Table 4-1 lists ail the test equipment required to do the
Performance Check Procedure. Test equipment specifi
cations described are the minimum necessary to pro
vide accurate results. For test equipment operating infor
mation, refer to the appropriate test equipment instruc
tion manual.
When you use equipment other than that recommended,
you may have to make some changes to the test setups.
If the exact example equipment in Table 4-1 is not avail
able, use the Minimum Specification column to deter
mine if any other available test equipment might be ade
quate to do the check.
PERFORMANCE CHECK INTERVAL
To ensure instrument accuracy, check the performance
of the 2246 1Y or 2246 Mod A after every 2000 hours of
operation, or once each year if used infrequently. If the
checks indicate a need for readjustment or repair, refer
the instrument to a qualified service person.
N O T E
The silver-oxide batteries used in the 2246 1Y
(instruments serial numbered B714999 and
below) MUST BE REPLACED AT LEAST
EVERY 12 MONTHS to ensure proper opera
tion of the instrument.
PREPARATION
This procedure is divided into subsections to let you
check individual sections of the instrument when it is not
necessary to do the complete Performance Check. An
Equipment Required block at the beginning of each sub
section lists the equipment from Table 4-1 that is needed
to do the checks in that subsection.
The initial front-panel control settings at the beginning of
each subsection prepare the instrument for the first step
of the subsection. Do each of the steps in a subsection
completely and in order to ensure the correct control set
tings for steps that follow. To ensure performance ac
curacies stated in the Specification (Section 1), let the in
strument warm up for 20 minutes and run the SELF CAL
MEASUREMENTS routine.
To run the SELF CAL MEASUREMENTS routine:
Press the top and bottom menu-item select buttons to
display the SERVICE MENU. Underline and select SELF
CAL MEASUREMENTS. Press RUN to start the routine,
then QUIT to return to the normal oscilloscope mode.
N O T E
Performance accuracies are ensured only
when the SELF CAL MEASUREMENTS are
done AFTER the 20-m inute warmup.
REV OCT 1991
4-1
Содержание 2246 1Y
Страница 13: ...2246A Service 7062 01 X The 2246 1Y or 2246 Mod A Portable Oscilloscope ...
Страница 35: ......
Страница 69: ...Theory of Operation 2246 1Y and 2246 Mod A Service 15 V 6081 07 3 32 Figure 3 4 Simplified Sweep Circuit ...
Страница 91: ...Theory of Operation 2246 1Y and 2246 Mod A Service 6081 12 6557 99 3 54 Figure 3 9 Power Supply block diagram ...
Страница 139: ......
Страница 185: ......
Страница 187: ......
Страница 278: ...Figure 9 5b Detailed 2246 1Y or 2246 Mod A block diagram part 2 R E ...
Страница 287: ......
Страница 304: ...A D D J U N 1991 ...
Страница 311: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 1 0 V r l III B f e I l i 6555 39 ...
Страница 314: ...15 14 13 12 11 10 9 8 7 6 5 12 11 10 9 8 7 AT117 A TI 271 1 2 3 4 5 6 6555 70 Figure 9 8 Hybrid pin identifiers ...
Страница 319: ......
Страница 320: ......
Страница 321: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 2 SET READOUT CONTROL CCW OFF 6555 40 ...
Страница 323: ......
Страница 324: ......
Страница 326: ...A B C P E F G ...
Страница 327: ......
Страница 331: ......
Страница 337: ......
Страница 340: ......
Страница 343: ...C l rti ...
Страница 344: ...WAVEFORMS FOR DIAGRAM 5 2246 1Y and 2246 Mod A Service HORIZ MODE A A SEC DIV 2M S HOLDOFF MIN CCW ...
Страница 345: ...HORIZ MODE ALT A SEC DIV 2M s B SEC DIV 5 mS HOLDOFF MIN DELAY INTENSIFIED ZONE STARTS MIDSCREEN ...
Страница 347: ...BEV JUNE 1991 7062 18 ...
Страница 349: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 6 6555 43 I ...
Страница 354: ...A L B i C 1 D 1 E i F i G 1 2 3 4 5 6 7 8 2246 1Y AND 2246 MOD A _________A fc A A REV JUNE 1991 7062 19 ...
Страница 355: ...I BOARD LOCATION 8K 8J 8H 7H 8H 8H 7J 8D 7H 8H 8J 8J 7K 8K 10D 6L 6K 6L 6K 6G 5H I ...
Страница 362: ......
Страница 372: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 9 MORE ...
Страница 373: ... 6555 72 6557 88 ...
Страница 375: ......
Страница 376: ...SCHEM LOCATION BOARD LOCATION 2E 2B 7F 4B 4J 5D 3J 5C 4L 6D 3L 6C 5L 7D 3L 7D 3M 7D 5M 7D IB 1A 1C 1A 2B 1A 3B 1A ...
Страница 380: ...WAVEFORMS FOR DIAGRAM 10 ...
Страница 387: ...WAVEFORMS FOR DIAGRAM 11 6555 73 6557 48 ...
Страница 394: ...WAVEFORMS FOR DIAGRAM 12 2246 1Y and 2246 Mod A Service i f I S 1 f c i 4 i ...
Страница 395: ...WAVEFORMS FOR DIAGRAM 12 cont 2246 1Y and 2246 Mod A Service ...
Страница 397: ......
Страница 398: ......
Страница 400: ...REV JUNE 1991 7062 39 ...
Страница 401: ......
Страница 403: ......
Страница 404: ......
Страница 406: ...B 2246 IV AND 2246 MOD A ...
Страница 407: ......
Страница 408: ......
Страница 414: ... _ 5 ...
Страница 415: ...2246 1Y 2246 MOD A ...
Страница 416: ......
Страница 423: ...2246 1Y 2246 MOD A ...
Страница 424: ...2246 1Y 2246 MOD A ...
Страница 425: ......
Страница 433: ......