Appendix D: Performance Tests
Appendices
AĆ64
d. Vertically center the display.
e. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 50 kHz.
f.
CHECK for a triggered display (
TRIG'D
indicator light on).
g. Set the
SEC/DIV
control to 50 ns.
h. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 10 MHz.
i.
CHECK for a triggered display (
TRIG'D
indicator light on).
j.
Disconnect the probe from the test equipment.
3. Check Motor Trigger
a. Set:CH 1
VOLTS/DIV
0.1 V
SEC/DIV
1 ms
Trigger
MODE
AUTO BL
Trigger
SOURCE
VERT
TRIG POS
MID
MOTOR TRIG
ON
b. Connect the fast rise output of the Calibration Generator
through the BNCĆtoĆbananaĆplug adapter to the CH 1 probe
tip and adjust the fast rise amplitude for 4 divisions.
c. Set the Calibration Generator frequency to 1 kHz.
d. Press the
AUTO:LVL PUSH
button to center the trigger level
on the waveform.
e. Adjust the Calibration Generator frequency variable to the
point where the trigger of the waveform is just able to be
maintained.
f.
Verify that the negative portion of the displayed square wave
is between 2.0 and 3.0 ms.
g. Disconnect the test equipment from the instrument.
Appendix D: Performance Tests
Appendices
AĆ64
d. Vertically center the display.
e. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 50 kHz.
f.
CHECK for a triggered display (
TRIG'D
indicator light on).
g. Set the
SEC/DIV
control to 50 ns.
h. Set the Leveled Sine Wave Generator to display 5 divisions of
amplitude at 10 MHz.
i.
CHECK for a triggered display (
TRIG'D
indicator light on).
j.
Disconnect the probe from the test equipment.
3. Check Motor Trigger
a. Set:CH 1
VOLTS/DIV
0.1 V
SEC/DIV
1 ms
Trigger
MODE
AUTO BL
Trigger
SOURCE
VERT
TRIG POS
MID
MOTOR TRIG
ON
b. Connect the fast rise output of the Calibration Generator
through the BNCĆtoĆbananaĆplug adapter to the CH 1 probe
tip and adjust the fast rise amplitude for 4 divisions.
c. Set the Calibration Generator frequency to 1 kHz.
d. Press the
AUTO:LVL PUSH
button to center the trigger level
on the waveform.
e. Adjust the Calibration Generator frequency variable to the
point where the trigger of the waveform is just able to be
maintained.
f.
Verify that the negative portion of the displayed square wave
is between 2.0 and 3.0 ms.
g. Disconnect the test equipment from the instrument.
Содержание 222PS
Страница 6: ......
Страница 14: ...Contents viii Glossary Index Glossary GĆ1 Index IĆ1 Contents viii Glossary Index Glossary GĆ1 Index IĆ1 ...
Страница 18: ...Contents xii Contents xii ...
Страница 21: ...Overview Overview ...
Страница 22: ......
Страница 27: ...At a Glance At a Glance ...
Страница 28: ......
Страница 45: ...In Detail In Detail ...
Страница 46: ......
Страница 50: ...Acquisition Modes In Detail 3Ć4 Acquisition Modes In Detail 3Ć4 ...
Страница 54: ...Auto Setup In Detail 3Ć8 Auto Setup In Detail 3Ć8 ...
Страница 60: ...Calibration In Detail 3Ć14 Calibration In Detail 3Ć14 ...
Страница 66: ...Channels In Detail 3Ć20 Channels In Detail 3Ć20 ...
Страница 70: ...The Display In Detail 3Ć24 The Display In Detail 3Ć24 ...
Страница 76: ...Horizontal Operation In Detail 3Ć30 Horizontal Operation In Detail 3Ć30 ...
Страница 96: ...Probes In Detail 3Ć50 Probes In Detail 3Ć50 ...
Страница 106: ...Saving and Recalling Data In Detail 3Ć60 Saving and Recalling Data In Detail 3Ć60 ...
Страница 108: ...Store Mode In Detail 3Ć62 Store Mode In Detail 3Ć62 ...
Страница 124: ...Vertical Operation In Detail 3Ć78 Vertical Operation In Detail 3Ć78 ...
Страница 127: ...Appendices Appendices ...
Страница 128: ......
Страница 164: ...Appendix B Remote Communication Appendices AĆ36 Appendix B Remote Communication Appendices AĆ36 ...
Страница 170: ...Appendix C Quick Checks Appendices AĆ42 Appendix C Quick Checks Appendices AĆ42 ...
Страница 210: ...Appendix F Accessories Appendices AĆ82 Appendix F Accessories Appendices AĆ82 ...
Страница 211: ...Glossary Index Glossary Index ...
Страница 212: ......
Страница 230: ...Index Index IĆ10 Index Index IĆ10 ...
Страница 231: ......
Страница 232: ......