222PS Operator Manual
3Ć9
Calibration
The 222PS has a vertical channel selfĆcalibration routine to maintain
best balance and accuracy with temperature variations.
Running the SelfĆCalibration Routine
To achieve the specified performance, you should recalibrate the
222PS any time the ambient temperature has changed by more than
5° C. If the trace jumps when you rotate the volts per division knob
with no signal applied, the instrument probably needs recalibration.
To run the selfĆcalibration routine, follow these steps.
ĂĂStep 1:ă
Disconnect both probes from the signal source.
NOTE
To ensure that the selfĆcalibration routine produces accuĆ
rate results, do not run the selfĆcalibration routine while
either probe is connected to a signal source.
ĂĂStep 2:ă
Press the
AUX FUNCT
button on the top panel to display
the auxiliary functions menu. The display appears as shown in
Figure 3Ć3.
222PS Operator Manual
3Ć9
Calibration
The 222PS has a vertical channel selfĆcalibration routine to maintain
best balance and accuracy with temperature variations.
Running the SelfĆCalibration Routine
To achieve the specified performance, you should recalibrate the
222PS any time the ambient temperature has changed by more than
5° C. If the trace jumps when you rotate the volts per division knob
with no signal applied, the instrument probably needs recalibration.
To run the selfĆcalibration routine, follow these steps.
ĂĂStep 1:ă
Disconnect both probes from the signal source.
NOTE
To ensure that the selfĆcalibration routine produces accuĆ
rate results, do not run the selfĆcalibration routine while
either probe is connected to a signal source.
ĂĂStep 2:ă
Press the
AUX FUNCT
button on the top panel to display
the auxiliary functions menu. The display appears as shown in
Figure 3Ć3.
Содержание 222PS
Страница 6: ......
Страница 14: ...Contents viii Glossary Index Glossary GĆ1 Index IĆ1 Contents viii Glossary Index Glossary GĆ1 Index IĆ1 ...
Страница 18: ...Contents xii Contents xii ...
Страница 21: ...Overview Overview ...
Страница 22: ......
Страница 27: ...At a Glance At a Glance ...
Страница 28: ......
Страница 45: ...In Detail In Detail ...
Страница 46: ......
Страница 50: ...Acquisition Modes In Detail 3Ć4 Acquisition Modes In Detail 3Ć4 ...
Страница 54: ...Auto Setup In Detail 3Ć8 Auto Setup In Detail 3Ć8 ...
Страница 60: ...Calibration In Detail 3Ć14 Calibration In Detail 3Ć14 ...
Страница 66: ...Channels In Detail 3Ć20 Channels In Detail 3Ć20 ...
Страница 70: ...The Display In Detail 3Ć24 The Display In Detail 3Ć24 ...
Страница 76: ...Horizontal Operation In Detail 3Ć30 Horizontal Operation In Detail 3Ć30 ...
Страница 96: ...Probes In Detail 3Ć50 Probes In Detail 3Ć50 ...
Страница 106: ...Saving and Recalling Data In Detail 3Ć60 Saving and Recalling Data In Detail 3Ć60 ...
Страница 108: ...Store Mode In Detail 3Ć62 Store Mode In Detail 3Ć62 ...
Страница 124: ...Vertical Operation In Detail 3Ć78 Vertical Operation In Detail 3Ć78 ...
Страница 127: ...Appendices Appendices ...
Страница 128: ......
Страница 164: ...Appendix B Remote Communication Appendices AĆ36 Appendix B Remote Communication Appendices AĆ36 ...
Страница 170: ...Appendix C Quick Checks Appendices AĆ42 Appendix C Quick Checks Appendices AĆ42 ...
Страница 210: ...Appendix F Accessories Appendices AĆ82 Appendix F Accessories Appendices AĆ82 ...
Страница 211: ...Glossary Index Glossary Index ...
Страница 212: ......
Страница 230: ...Index Index IĆ10 Index Index IĆ10 ...
Страница 231: ......
Страница 232: ......