Calibration
222PS Operator Manual
3Ć11
ĂĂStep 4:ă
To begin the selfĆcalibration routine for channel 1, press
the menu button next to
CH1
.
This routine takes about a minute to perform. When it is finished,
the instrument displays PASS or FAIL.
If PASS appears on the screen, channel 1's recalibration was
successful.
If FAIL appears on the screen, run the calibration routine again. If
the failure persists, refer the instrument to a qualified service
person.
If the selfĆcalibration routine fails, the instrument sends a coded
error message. To capture this message, connect the instrument
to a terminal (or a PC emulating a terminal) through the RSĆ232
port on the rear panel and run the failed routine again. See
Appendix B: Remote Communications
for a discussion of the error
codes.
ĂĂStep 5:ă
Now calibrate channel 2. Invoke the menu again, and
press the menu button next to
CH2
. Follow the procedure outlined
in Step 4.
ĂĂStep 6:ă
Now recalibrate the external trigger input. Before you start
the calibration, connect the external trigger input connector to the
trigger common reference connector on the rear panel. Use a
jumper cable with a banana plug connector on each end.
Calibration
222PS Operator Manual
3Ć11
ĂĂStep 4:ă
To begin the selfĆcalibration routine for channel 1, press
the menu button next to
CH1
.
This routine takes about a minute to perform. When it is finished,
the instrument displays PASS or FAIL.
If PASS appears on the screen, channel 1's recalibration was
successful.
If FAIL appears on the screen, run the calibration routine again. If
the failure persists, refer the instrument to a qualified service
person.
If the selfĆcalibration routine fails, the instrument sends a coded
error message. To capture this message, connect the instrument
to a terminal (or a PC emulating a terminal) through the RSĆ232
port on the rear panel and run the failed routine again. See
Appendix B: Remote Communications
for a discussion of the error
codes.
ĂĂStep 5:ă
Now calibrate channel 2. Invoke the menu again, and
press the menu button next to
CH2
. Follow the procedure outlined
in Step 4.
ĂĂStep 6:ă
Now recalibrate the external trigger input. Before you start
the calibration, connect the external trigger input connector to the
trigger common reference connector on the rear panel. Use a
jumper cable with a banana plug connector on each end.
Содержание 222PS
Страница 6: ......
Страница 14: ...Contents viii Glossary Index Glossary GĆ1 Index IĆ1 Contents viii Glossary Index Glossary GĆ1 Index IĆ1 ...
Страница 18: ...Contents xii Contents xii ...
Страница 21: ...Overview Overview ...
Страница 22: ......
Страница 27: ...At a Glance At a Glance ...
Страница 28: ......
Страница 45: ...In Detail In Detail ...
Страница 46: ......
Страница 50: ...Acquisition Modes In Detail 3Ć4 Acquisition Modes In Detail 3Ć4 ...
Страница 54: ...Auto Setup In Detail 3Ć8 Auto Setup In Detail 3Ć8 ...
Страница 60: ...Calibration In Detail 3Ć14 Calibration In Detail 3Ć14 ...
Страница 66: ...Channels In Detail 3Ć20 Channels In Detail 3Ć20 ...
Страница 70: ...The Display In Detail 3Ć24 The Display In Detail 3Ć24 ...
Страница 76: ...Horizontal Operation In Detail 3Ć30 Horizontal Operation In Detail 3Ć30 ...
Страница 96: ...Probes In Detail 3Ć50 Probes In Detail 3Ć50 ...
Страница 106: ...Saving and Recalling Data In Detail 3Ć60 Saving and Recalling Data In Detail 3Ć60 ...
Страница 108: ...Store Mode In Detail 3Ć62 Store Mode In Detail 3Ć62 ...
Страница 124: ...Vertical Operation In Detail 3Ć78 Vertical Operation In Detail 3Ć78 ...
Страница 127: ...Appendices Appendices ...
Страница 128: ......
Страница 164: ...Appendix B Remote Communication Appendices AĆ36 Appendix B Remote Communication Appendices AĆ36 ...
Страница 170: ...Appendix C Quick Checks Appendices AĆ42 Appendix C Quick Checks Appendices AĆ42 ...
Страница 210: ...Appendix F Accessories Appendices AĆ82 Appendix F Accessories Appendices AĆ82 ...
Страница 211: ...Glossary Index Glossary Index ...
Страница 212: ......
Страница 230: ...Index Index IĆ10 Index Index IĆ10 ...
Страница 231: ......
Страница 232: ......