Power
In Detail
3Ć42
ĂĂStep 2:ă
Plug the prong end of the External Power AC Adapter into
an AC power source.
ĂĂStep 3:ă
Press the
ON
button.
When the oscilloscope is operating under external power, an
external power indicator
appears in the upper right corner of
the display.
Other Sources of External Power
You can operate the 222PS from your own external power source. The
power source must supply at least 15ĂW or 16ĂvoltĆamperes.
H
An AC power source must provide 16-20ĂVAC at 47-400 ĂHz.
H
A DC power source must provide 12-28ĂVDC.
The external power input connector has two contacts. DC power of
either polarity can be between contacts.
NOTE
To prevent blowing internal fuses, do not force either pin
lower than .5 volts more negative than the instrument
chassis. The instrument chassis is connected to the
ground pin of the RSĆ232 communications port and to the
external trigger connector common.
WARNING
To avoid possible injuryor damage to the 222PS or
equipment connected to it, do not float the external trigger
common connector, the RSĆ232 communications port, or
the external power input above 42ĂV peak. These inputs
are not electricallyisolated from each other.
Power
In Detail
3Ć42
ĂĂStep 2:ă
Plug the prong end of the External Power AC Adapter into
an AC power source.
ĂĂStep 3:ă
Press the
ON
button.
When the oscilloscope is operating under external power, an
external power indicator
appears in the upper right corner of
the display.
Other Sources of External Power
You can operate the 222PS from your own external power source. The
power source must supply at least 15ĂW or 16ĂvoltĆamperes.
H
An AC power source must provide 16-20ĂVAC at 47-400 ĂHz.
H
A DC power source must provide 12-28ĂVDC.
The external power input connector has two contacts. DC power of
either polarity can be between contacts.
NOTE
To prevent blowing internal fuses, do not force either pin
lower than .5 volts more negative than the instrument
chassis. The instrument chassis is connected to the
ground pin of the RSĆ232 communications port and to the
external trigger connector common.
WARNING
To avoid possible injuryor damage to the 222PS or
equipment connected to it, do not float the external trigger
common connector, the RSĆ232 communications port, or
the external power input above 42ĂV peak. These inputs
are not electricallyisolated from each other.
Содержание 222PS
Страница 6: ......
Страница 14: ...Contents viii Glossary Index Glossary GĆ1 Index IĆ1 Contents viii Glossary Index Glossary GĆ1 Index IĆ1 ...
Страница 18: ...Contents xii Contents xii ...
Страница 21: ...Overview Overview ...
Страница 22: ......
Страница 27: ...At a Glance At a Glance ...
Страница 28: ......
Страница 45: ...In Detail In Detail ...
Страница 46: ......
Страница 50: ...Acquisition Modes In Detail 3Ć4 Acquisition Modes In Detail 3Ć4 ...
Страница 54: ...Auto Setup In Detail 3Ć8 Auto Setup In Detail 3Ć8 ...
Страница 60: ...Calibration In Detail 3Ć14 Calibration In Detail 3Ć14 ...
Страница 66: ...Channels In Detail 3Ć20 Channels In Detail 3Ć20 ...
Страница 70: ...The Display In Detail 3Ć24 The Display In Detail 3Ć24 ...
Страница 76: ...Horizontal Operation In Detail 3Ć30 Horizontal Operation In Detail 3Ć30 ...
Страница 96: ...Probes In Detail 3Ć50 Probes In Detail 3Ć50 ...
Страница 106: ...Saving and Recalling Data In Detail 3Ć60 Saving and Recalling Data In Detail 3Ć60 ...
Страница 108: ...Store Mode In Detail 3Ć62 Store Mode In Detail 3Ć62 ...
Страница 124: ...Vertical Operation In Detail 3Ć78 Vertical Operation In Detail 3Ć78 ...
Страница 127: ...Appendices Appendices ...
Страница 128: ......
Страница 164: ...Appendix B Remote Communication Appendices AĆ36 Appendix B Remote Communication Appendices AĆ36 ...
Страница 170: ...Appendix C Quick Checks Appendices AĆ42 Appendix C Quick Checks Appendices AĆ42 ...
Страница 210: ...Appendix F Accessories Appendices AĆ82 Appendix F Accessories Appendices AĆ82 ...
Страница 211: ...Glossary Index Glossary Index ...
Страница 212: ......
Страница 230: ...Index Index IĆ10 Index Index IĆ10 ...
Страница 231: ......
Страница 232: ......