Calibration— Type 1A1
NOTE
All waveforms shov/n in this procedure are actual
waveform photographs taken with a Tektronix
Oscilloscope Camera system. Projected graticules
were used. Each major division on the graticule
illustrations represents one cm.
Preliminary Procedure
1. Remove the panels from the test oscilloscope to expose
the left side and bottom of the vertical plug-in compart
ment.
2. Lay the oscilloscope on its right side for access to the
bottom side of the Type 1A1.
3. Install the Type 1A1 in the oscilloscope vertical plug
in compartment.
4. Connect the power cord of the oscilloscope to the
design-center operating voltage for which the oscilloscope
is wired.
5. Turn on the oscilloscope and allow 15 minutes for warm
up and stabilization. While the oscilloscope is warming up,
preset the controls to the position given in the list that pre
cedes step
1
.
7-5
Содержание 1A1
Страница 47: ...Maintenance Type 1A1 670 0075 01 Fig 5 5 Channel 1 Input Amplifier card 5 13 http manoman sqhill com ...
Страница 50: ...Maintenance Type 1AT Fig 5 8 Front of Output Amplifier card 5 16 http manoman sqhill com ...
Страница 128: ...ALT SWEEP SLAVE PULSE CHOPPED BLANKING ALT TRACE SYNC PULSE N 2 0 0 0 0 U P SWITCHING CIRCUIT TYPE IAI PLUG IN ...
Страница 129: ...http manoman sqhill com ...
Страница 130: ...38 ...
Страница 131: ...FIG 3 CIRCUIT CARDS ACCESSORIES CHANNEL I INPUT OUTPUT AMPLIFIER CHANNEL 2 INPUT TYPE 1A1 DUAL TRACE PLUG IN UNIT ...
Страница 134: ...Page 2 o f 2 TYPE 1A1 A p a r t i a l CHAKJKJEL Z INPUT A M P L I F I E R M16 025 1269 ...